Chapter 4: Parameters Setting and Preventative Maintenance
28
7.
Instrument Door and Ceramic Disks
The instrument door must fit properly for the system to operate correctly. The platen part
of the roller pump is located on the door. The platen must line up properly with the pump.
a.
Check hinges for blood build-up, clean any dried blood from hinge area. Be sure
that door is seated completely down on the hinges.
b.
Check plastic rivets and door integrity. Make sure that the door frame is not bent.
Replace, if bent.
c.
Inspect the ceramic disks on the door and in the center of the unit for cracks.
Return to manufacturer for replacement if they are damaged.
8.
Rubber Feet
Inspect the rubber feet on the bottom of the unit for cracked or missing rubber feet.
Replace if necessary.