Chapter 3: Alarms and Troubleshooting Guide
16
ALARM MESSAGE
POSSIBLE
CONDITION
OPERATOR ACTION
HIGH PRESSURE
High Pressure Alarm Screen
Patient line is
blocked.
Recirculate line is
blocked.
Infusion site is not
well placed.
The catheter bore
size is too small.
Pressure limit
setting is set too
low.
Make certain that the flow path is not
blocked.
Check that the recirculate line is not
obstructed.
Check that the infusion site is well
placed and use the appropriate
infusion set recommended in the
guide, Match the Infusion Set to Flow
Rate and Fluid Type on page 10.
Increase pressure limit setting.
LOW BATTERY
Plug the system into an AC outlet to
continue operation and recharge the
battery. Allow at least 8 hours to fully
charge the battery.
Battery voltage is
too low
If LOW BATTERY displayed while the
system is connected to AC power, one
of the components may be defective.
Service machine.
If battery is completely discharged,
turn the AC power OFF, plug the
system into an AC outlet to recharge
the battery.
Wait for at least 30
seconds before turning the system
ON.
MISSING DISPOSABLE
Missing Disposable Screen
No disposable set
in the unit.
Properly install disposable.
Press NEXT to resume.