Chapter 4: Parameters Setting and Preventative Maintenance
42
Table 201
Guidance and Manufacturer’s Declaration – Emissions All Equipment and Systems
The
Belmont
®
Rapid Infuser, FMS2000
is intended for use in the electromagnetic environment specified below. The
customer or user of the
Belmont
®
Rapid Infuser, FMS2000
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Enforcement
– guidance
RF Emissions
CISPR 11
Group 1, Class A
The
Belmont
®
Rapid Infuser, FMS2000
uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Harmonics IEC
61000-3-2
Complies or Not
applicable
Complies
Flicker
IEC 61000-3-3
Complies or Not
applicable
Complies
Table 202
Guidance and Manufacturer’s Declaration—Immunity All Equipment and Systems
The
Belmont
®
Rapid Infuser, FMS2000
is intended for use in the electromagnetic environment specified below. The
customer or user of the
Belmont
®
Rapid Infuser, FMS2000
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6kV contact
±8kV air
±6kV contact
±8kV air
Floors should be wood, concrete or
ceramic tile. If floors are synthetic, the
relative humidity should be at least
30%
.
Radiated RF
IEC 61000-4-3
3 V/m
3 V/m
Radiated RF frequency is negligible at
any single frequency
Electrical Fast
Transient/burst
IEC 61000-4-4
±2kV on AC Mains
±2kV on AC Mains
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
±1kV Differential
±2kV Common
±1kV Differential
±2kV Common
Mains power quality should be that of
a typical commercial or hospital
environment.
Conducted RF
IEC 61000-4-6
3 Vrms from
150 kHz to 80 MHz
V1 = 3 Vrms
If interference occurs, it may be
necessary to position the
Belmont
®
Rapid Infuser, FMS2000
further from
sources of power frequency magnetic
field.
Power Frequency
50/60Hz Magnetic
Field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields
should be that of a typical location in a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for5 Seconds
>95% Dip for 0.5 Cycle
60% Dip for 5 Cycles
30% Dip for 25 Cycles
>95% Dip for 5 Seconds
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
Belmont
®
Rapid Infuser, FMS2000
requires continued operation during
power mains interruptions, it is
recommended that the
Belmont
®
Rapid Infuser, FMS2000
be powered
from an uninterruptible power supply
or battery.