Chapter 4: Parameters Setting and Preventative Maintenance
38
6.
Clean Pump Head
The pump head can be removed and cleaned if needed.
a.
Turn the pump to STANDBY and unplug the power cord.
b.
Unscrew the retaining screw that holds the pump head.
c.
Remove the pump head and clean with water and soap. Hydrogen peroxide or a
mild bleach solution can be used to disinfect.
d.
Let pump head dry before replacing and make certain the pump head is securely
fastened with the retaining screw.
e.
If the pump head squeaks, spray the roller with Silicone spray (Heavy Duty Pure
Silicone.)
Retaining Screw
Roller Pump