Chapter 2: Operation
8
POWER
ON
Check that the detachable power cable
is securely seated in the main power
receptacle.
Plug the system power cord into a
grounded, 3-prong, 20 Amp, AC
receptacle. Do not use an adaptor for
ungrounded outlets.
Power-Up Screen
U
Missing Disposable Screen
1.
Turn power on by firmly pressing the circuit
breaker to the ON position. The system will
perform a self-check to check the integrity of
system parameters.
2.
AC POWER PRESENT appears at the logo
screen when the system first powers up.
Check the power cord and AC receptacle
connections if the statement does not appear.
3.
PRIME screen will appear.
4.
Press NEXT to go to the PRIME screen.
If you turn power ON without the
disposable set, MISSING DISPOSABLE
audible alarm message will appear.
Open the door or press MUTE to silence
the alarm then install the disposable set.
INSTALLING FLUID BAG
Install solution compatible with blood for
the main system prime.
,
Prime Screen
1.
Hang fluid bag(s) on the IV pole.
2.
Completely close bag clamps, remove the bag
spike cap(s). Spike fluid bag(s), pierce it fully
to ensure that fluids flow freely.
3.
Open bag clamps.
When hanging the fluid bag above the
machine, the pump tubing that is seated in
the fluid out detector should not be
stretched. Stretching the pump tubing may
cause false Fluid Out alarms.
The recirculate line must not be kinked or
restricted.