Chapter 5: Technical Specifications
44
Technical Specifications of the Belmont
FMS2000
DIMENSION
Size
13.5" x 12" x 7.5"
(34.29cm x 30.48cm x 19.05cm)
Weight
28 lbs (12.7 Kg)
PORTABILITY
Hand Carry
Handle on top of unit for easy transport
IV Pole Mount
IV pole mountable or free standing. I.V pole diameter
range of pole mount: 1" - 1 1/4"
POWER
AC
Input Voltage
115-120 V ~ 20 amp dedicated
or 230 V ~ 16 amp dedicated
Fuse
1.25A, 250V, Fast Acting, 5x20mm
Operating Frequency
50/60 Hz
Maximum Power
1440 VA
Line Isolation
1500 V to ground
Earth Leakage Current
< 300 µA (For Domestic unit)
< 500 µA (For 230 V ~ unit)
Electrical
Compliance
EN 60601-1, CSA/C22.2 - No. 601.1-M90
Circuit Breaker
15Amp, 125VAC/250VAC, 50/60 Hz
Power Cord
U.S.: 3 conductors, 14 AWG type SJT Cord with
Hospital grade plug
Outside U.S.: 3 x 1.5 mm
2
International Harmonized
Cordage with Hospital grade plug
Battery Type
Running Time
Recharge Time
Rechargeable lead acid
> 30 minutes at 50ml/min. without heat
8 hours