OVERVIEW OF THE BELMONT
FMS2000
The complete system consists of the FMS2000
Control System
, which is mounted on an IV pole,
and the FMS2000
Disposable Set
.
The
FMS2000
can be used only with the supplied
disposables.
A large volume 3-liter reservoir is available as an optional accessory for
convenience in cases involving very large infusion volumes, see page 7.
The
Disposable Set
is preconnected and has a sterile fluid path.
It is intended for single
patient use only
.
System Diagram Showing Main Components
Major components of the Control System
:
1. Fluid Out Detector detects and alarms at
an out of fluid situation.
2. Power Switch turns system power on and
off.
3. Roller Pump is designed for pumping
accuracy and reliability.
4. Display and control panel show status and
alarm messages with touch keys at the
bottom of the screen.
5. IR Temperature Probe (Output Probe)
monitors output fluid temperature as it
exits the Heat Exchanger.
6. Air Detector detects air in the line. If air is
detected the valve wand is closed
immediately to prevent air into the patient.
Pumping and heating stop, alarm sounds
and “Air Detection” message is displayed
on screen.
7. Valve wand closes off the recirculation
line when the system is in the infusion
mode and closes off the infusion line
when the system is in the recirculation
mode. It immediately closes the infusion
line to the patient when an error condition
occurs
which
may
require
user
intervention.
8. IR Temperature Probe (Input Probe)
monitors input fluid temperature as it
enters the Heat Exchanger.