8
microspeed® uni motor system
Aesculap Power Systems
microspeed® uni motor cable for foot control GD672
Fig. 4
Legend
30
Motor release switch
Note
This motor cable can only be used in connection with a foot control.
The motor coupled to this motor cable can only be activated if motor
release switch
30
is in the “ON” position. When two motor cables GD672
are used at the same time, a motor can be activated only if it has been
released (motor release switch
30
in position "ON") while the other motor
is deactivated (motor release switch
30
in position "OFF").
microspeed® uni motor cable with hand control GD673
Fig. 5
Legend
25
Motor release switch:
Position "ON": Motor released for operation
Position "OFF": Motor locked against operation
26
Hand control function button:
Short actuation: Switching on/off the coolant pump
Pressing for more than three seconds (with the motor inactive):
Change to Device settings menu, siehe Kapitel Changing the settings
in the Device settings menu through hand control
29
Lever:
Actuator for motor speed setting
Note
This motor cable can not be used in connection with a foot control.
The motor coupled to this motor cable can only be activated if the motor
release switch is in its “ON” position.
30
26
29
25