microspeed® uni motor system
Aesculap Power Systems
A motor control field is displayed as an active control field when the motor
assigned to the motor control field:
• is operated
• was the last motor operated.
The symbol of the connected motor type is highlighted by a green
background whenever a motor control field is released for operation.
2.4.5 Pump control field
Fig. 3
Legend
A
Status indicator symbol of coolant pump
4
. The symbol is highlighted
by a green background while the pump is active. The pump can only
be activated if it has been released for operation and if the motor is
active.
B
Control keys for increasing or reducing the delivery rate
C
Display of selected dosage
The preset dose level
C
applies to the motor whose settings are displayed
in the (white background) motor control field.
2.4.6 Instruction menu
Note
The Instruction menu can only be called up at the control unit.
This control key is pressed to call up the Instruction menu. In the
Instruction menu, information about the error conditions output to the
display and device settings such as the user language and the volume of
the acoustic alarm/notification signal can be changed.
Motor-type specific adjustments and settings can be carried out here, too:
• Slow-down and speed-up ramps for each motor type
• Oscillating or tapping mode (for pistol handpiece GD684)
• Oscillation frequency and speed-up ramp in oscillating mode (in
Shaving mode of operation of shaver handpiece GD682)
• Reset: to restore the factory settings
2.4.7 Motor recognition
Control unit
1
recognizes the motor type connected to one of the motor
connection sockets
5
of control unit
1
and displays the type in the
respective motor control field of display
2
. The most recent settings for this
motor type at the same motor connection socket are shown on display
2
.
The settings can be changed within the motor-specific setting ranges,
either on the display or through hand or foot control (in the Device
settings menu), see chapter Technical specifications.
The motor control field of the most recently activated motor is always
brightly highlighted, whereas the motor control field of the motor that is
not the one that was most recently activated or locked is displayed with
a gray background. When only one motor is connected to control unit
1
,
the corresponding control field is highlighted, if this motor is not locked.
Saw program (only for low-speed motors GD678 and GD679)
In this program the maximum motor speed is limited to 16,000 1/min. The
settings most recently used with the saw program (motor speed range
upper limit, dose level for coolant pump
4
) are recalled automatically.
2.4.8 Automatic saving of device settings
The most recent settings for this motor type at this motor connection
socket (speed range upper limit, delivery rate, etc.), are called up
automatically as soon as the same motor type is recognized again as
connected at the same motor connection socket of control unit
1
after a
motor change.
Note
After each device start-up both the coolant pump and motor type-specific
special functions such as the saw program (for GD678 and GD679) or the
oscillating or tapping mode (for GD684) are always inactive (even if they
had been released or activated before). All parameters are reset when the
operating key "Factory setting" in the Instruction menu is pressed.