26
microspeed® uni motor system
Aesculap Power Systems
4.3.2 Setting up and operating the microspeed® uni shaver
handpiece GD682
Note
Shaver handpiece GD682 can only be used with foot control GD671 and a
surgical suction device (e.g. GF200 or GF210).
The instruction menu allows changing motor type-specific settings (slow-
down and speed-up ramps and oscillation frequency, speed-up ramp for
Shaving mode), see chapter Instruction menu.
Parameter settings can only be changed if:
• a motor with motor cable GD673 is connected at the second motor
connection socket without being active, or
• a pistol handpiece GD684 is connected at the second motor connection
socket without being active, or
• a motor with motor cable GD672 is connected at the second motor
connection socket and the motor release switch is in its “OFF” position,
or
• only one shaver handpiece GD682 is connected at the control unit and
• a foot control device is connected to the control unit.
The following settings cannot be changed while shaver handpiece GD682
is running:
• Motor direction (of rotation)
• Motor speed range upper limit
• Operating mode
All settings will be saved on exit from the settings mode.
Operating modes
There are two types of blades for shaver handpiece GD682:
• Reamer blades
• Shaver blades
Control unit
1
recognizes whether the coupled blade is a reamer or a
shaver blade. The respective mode of operation (reaming or shaving) is
indicated on the display.
An acoustic signal is issued whenever a blade of a different type is
inserted, and the respective symbol is displayed in the appropriate motor
control field.
With a reamer blade coupled, the motor can be operated in right or left
rotation, but not in oscillating mode. The maximum speed is limited to
8,000 1/min. The symbol for the "Reaming" operating mode is displayed in
the control field.
With a shaver blade coupled, the motor can be operated in right or left
rotation as well as in oscillating mode. The maximum speed is limited to
5,000 1/min. The symbol for the "Shaving" operating mode is displayed in
the control field.
Changing the settings in the Device settings menu on the control unit
Notes
For shaver handpiece GD682, the tool speed is indicated in the display.
The setting range and the step width for the speed range upper limit depend
on which operating mode (cutting or shaving) is selected, see chapter
Technical specifications.
Selecting the motor speed range upper limit
The set motor speed upper limit is indicated in the respective motor
control field in display
2
and can be adjusted via the touch control keys.
¾
To increase the motor speed range upper limit: press key to increase the
motor speed range upper limit.
¾
To reduce the motor speed range upper limit: press key to reduce the
motor speed range upper limit.
Setting the coolant pump delivery rate
Note
The coolant pump serves exclusively for cooling or irrigating the tools. The
delivery rate at the highest dose level is approx. 80 ml per minute.
Coolant pump
4
is activated and deactivated via the control key in the
respective motor control field.
The same control key also serves as a status indicator:
• The symbol displayed on a white background means that the pump is
switched off.
• The symbol displayed on a green background means that the pump is
switched on.
¾
To increase the delivery rate: press the key to increase the delivery rate
in the pump control field.
¾
To reduce the delivery rate: press the key to reduce the delivery rate in
the pump control field.
Changing the settings in the Device settings menu through foot
control GD671
¾
Press and hold down foot control function button
18
for three seconds.
Control unit
1
switches to the settings mode.
Shaver handpiece
37
and coolant pump
4
are disabled.
WARNING
Risk of injury caused by inadvertent activation of the
motor!
¾
Secure any motors that are not actively operated
against inadvertent activation through foot or hand
control.
WARNING
Risk of injury and material damage due to inappropriate use
of the blades!
¾
Always follow the safety advice and information given in
instructions for use TA011579.
¾
Do not exceed the specified maximum motor speed.