Horizon Nxt®
Modular Infusion System 8 - 8 6/00
950520 Rev F
Specifications
Mechanism:
Volumetric (Non Drop-Counting), Positive
Pressure Displacement Reservoir
Dimensions:
Width 12.4” (31.5 cm)
Height 4.8 (12.2 cm)
Depth 12.2” (31.5 cm)
Weight:
Approximately 11.9 lb. (5.4 Kg)
Power Requirements:
90-132 VAC, 30 Watts, 47/63 Hz excluding
accessories (domestic), 180- 264 VAC, 30
Watts, 47/63 Hz excluding accessories (Inter-
national)
Grounding Resistance
Meets UL Standard 544
Leakage Current:
Meets or exceeds UL Standard 544, Section
27.2, and ANSI/AAMI Safe Current Limits
(July 9, 1985)
Plug:
Hospital Grade (3 pin)
Pole Size Range:
0.75" to 1.25"
Battery:
12 Volt sealed lead-acid battery for ambulatory
use. Run time is dependent on the state of bat-
tery charge and the rate of fluid delivery.
Rate Range:
Standard: (Primary and Piggyback) 0.1 to
999.9 mL/hr
Micro: (Primary and Piggyback) 0.1 to 99.9
mL/hr
Volume to be Infused Range:
0.1 to 9999.9 mL
Occlusion Pressure
75, 100, 200, 300, 400, 500 and 750 mmHg
Standard mode default is 300 mmHg.
Keep Open Rate:
3 mL/hr for prescribed delivery rates of 3 to
999.9 mL/hr, or the continuation of the pre-
scribed delivery rate for 0.1 to 2.9 mL/hr set-
tings.
Volume Delivered Accuracy:
+/- 5%
Fluid Types:
All standard I.V. fluids