Horizon Nxt®
Modular Infusion system 4 - 1 6/00
950520 Rev F
Chapter 4
MENUS
The Horizon Nxt® can be customized to fit your specific needs. The User
Menu and Alternate Menu contain options identified by each individual
hospital. The following options will always appear in one of the 2 menus.
Main Menu
Change Alarm Volume changes the volume of the audible alarm.
Clear Total Infused clears the total IV fluid infused shown on the
Information Screen.
Dose Mode calculates the rate or dose information necessary for use with
many critical drug infusions.
Occlusion Limit sets the occlusion pressure limit to 75, 100, 200, 300, 400
or 500 mmHg.
Alternate Menu lists options used less often by the user.
Exit from Menu returns to the Information Screen.
Alternate Menu
Calculate BSA calculates the body surface area (BSA).
Set LCD Contrast changes the readability of the Information Screen.
Biomed Options allows access to the Biomed Menu.
Exit from Menu returns to the Information Screen.