Horizon Nxt®
Modular Infusion System 8 - 3 6/00
950520 Rev F
Alarm Troubleshooting Guide
MESSAGE
LED MESSAGES
CORRECTION
Air-in-Line
Air
Air in downstream tubing
Tubing improperly inserted in Air-
in-Line Detector.
Battery Very Low
Use AC
Plug into AC power NOW to contin-
ue operation
Container Empty
SOLU
Check for air in set
IV container empty
Check upstream clamp
Check cassette installation
Door Open
door
The door must be shut to deliver flu-
id
Downstream Occlusion
OCCL
Clamp closed or filter blocked
Tubing kinked
IV positional or infiltrated
Catheter or vein too small
Check occlusion limit
Hold Time Exceeded
HOLd
Press Hold key, three minute time
period is restarted
KVO
flashing LED’s
Pump is operating in Keep Vein
Open state
Infusion is complete
Set volume to be delivered
Restart or turn off the pump
Low Battery
PLUG IN
Battery too low; plug in AC
System Error
SYS Err
Turn the pump off and return to Bio-
medical Professional for service
See Help
SEE HELP
CLOSE ROLLER CLAMP, then
open door. Press LOAD and reload
tubing.
Cassette not in place
Set Piggyback Rate
Set a Piggyback rate