Horizon Nxt®
Modular Infusion System 2 - 4 6/00
950520 Rev F
Important Keys
An understanding of the function and location of these keys will assist in the
use of the pump.
From Figure 1:
• Power key (
6
) turns the pump on and off
• Data keys (
11
) allow input of data values
• Hold key (
12
) silences an alarm, stops an infusion, is a quick escape from
the menu, and extends the Hold state an additional 3 minutes every time it
is pushed if Hold Extender is active.
• Menu Key (
13
) lets the user see options using a menu screen
• Cursor keys (
14)
move the cursor bar on the Information Screen
• Enter key (
15
) validates or accepts the selection made
From Figure 2:
• Panel Lock Out Switch locks the keypanel to prevent tampering. The
switch can only be used while the pump is infusing.
Daisy Chaining
Daisy chaining allows up to 5 Horizon Nxt® or Horizon® pumps to be
plugged into each other with only one pump plugged into a wall outlet. On
the side of each Horizon Nxt pump is an electrical outlet. Only another
Horizon or Horizon Nxt pump should be plugged into that outlet.
Caution:
To daisy chain more than 5 pumps, verify that the total leakage
current as a system does not exceed 100 uA.
Battery Operation
The Horizon Nxt has been equipped with an internal, rechargeable battery
which will power the pump automatically if the plug is disconnected during
use. Battery power is provided as a back-up system only, and the length of
the battery life is affected by the rate of delivery. Whenever possible, the
pump should remain plugged into an electrical outlet. The Battery Use
indicator will be lit when the pump is operating on its battery.
The pump must remain plugged into a hospital-grade electrical outlet for at
least 24 hours to fully recharge the battery from a totally discharged
condition.