Horizon Nxt®
Modular Infusion System 7 - 5 6/00
950520 Rev F
NOT ALLOWED
ALLOWED
1) Press
R
UN
key to start the infusion
2) Period 1 is infusing as displayed
3) Press
E
NTER
key to display a graph
4) Press
E
NTER
key to return to the text display
Each period is represented by a vertical bar which fills as the fluid is infused.
The height of the bar represents the rate of infusion in that particular period.
The period presently being infused is displayed in the upper left corner. If
the vertical bar appears to be missing, the rate in that period is much less
compared to the other periods.
Review/Change PROGRAM:
Note:
A Program cannot be changed after it has been started.
1) Press
M
ENU
key
2) Press
C
URSOR
keys to select “Review/Change PROGRAM”
3) Press
E
NTER
key
1) Press
C
URSOR
keys to select rate
2) Press
D
ATA
keys to change the rate values
3) Press
E
NTER
key
Note:
Only rate can be changed after an infusion has started
4) Press
C
URSOR
keys to select “Validate”
5) Press
E
NTER
key
Restart PROGRAM:
1 50.0 35.0
2 35.0 0.0
3 30.0 500.0 ZERO
VALIDATE TOTAL = 0.0 ml
Pd. Rate Volum
e
Pd. Rate Volum
e
1 50.0 35.0
2 100.0 500.0
3 0.0 0.0 ZERO
VALIDATE TOTAL = 0.0 ml