Horizon Nxt® with MedScan
Modular Infusion System 5 - 7 6/00
950520 Rev F
midnight and 12:00 to represent noon. Verify that the Time
is correct.
1) Press
C
URSOR
to select "Start Dose at"
2) Press
D
ATA
keys to set the time the infusion is to begin
3) Press the
C
URSOR
keys to select "Validate"
4) Press
E
NTER
key
Note:
"
q(x)
Schedule Overlap"
message appears when doses
entered are not at least 15 minutes
apart. When this occurs the
q(x)°
feature must be exited to cancel the
present infusion regimen.
"Wait Is" shows the amount of time
until delivery starts
Note:
After pressing Enter, the pump goes into a waiting period.
"SCHd dLY" shows on the LED’s. No fluid is infusing
during this period.
1) Press
E
NTER
key for the data screen
2) Follow the steps above to change the data, if necessary
3) Press
E
NTER
key
At the
q
(x)° Start time, the pump will begin the infusion. During the
infusion "SCHd run" shows on the LED’s. The number of doses
remaining is shown on the LCD. After the infusion is complete, the
pump will automatically return to the waiting period. The pump will
continue to "infuse and wait" for the total number of doses entered.
To Exit:
q(x) Schedule Overlap
Press HOLD to Silence Alarm
q(x) Start 10:35a DATA
Wait Is 0 h 49 m
3 Doses left
Total Infused = 2.2 ml
q(x) Delivery DATA
Time Left = 0 hr 29 min
Total Infused = 2.7 ml
Occlusion Limit = 300 mmHg
3 Doses left