HorizonNxt®
Modular Infusion System 6 - 4 6/00
950520 Rev F
1) Press the left or right
C
URSOR
keys to
move to the appropriate parameter
2) Press the up or down
C
URSOR
keys to
change to the correct parameters (mcg/
kg/min to units/hr)
3) Press
E
NTER
key
Titrate Rate:
The rate can be titrated while the pump is running. Press the
D
ATA
keys to
change the rate. The dose is recalculated and shown. The
E
NTER
key or
R
UN
key must be pressed within four seconds to validate the change. Once made,
the pump infuses at the new rate.
Note
: If the
E
NTER
key or
R
UN
key is not pressed within 4 seconds of
entering a new rate, a series of error beeps will occur and the rate
change will be cancelled. The pump will then display the message
"Rate change aborted", and will continue to infuse at the old rate.
An intermittent alarm will continue to sound to notify the user that a
rate change has been attempted and cancelled. Pressing any key will
silence the alarm.
Titrate Dose:
1) Press
E
NTER
key
2) Press
C
URSOR
keys to select the Dose value
3) Press
D
ATA
keys to change the values
4) Press
C
URSOR
keys to select "
VALIDATE
"
5) Press
E
NTER
key
Monitor the infusion as indicated by hospital policy. The pump will alarm
and go into the KVO state when the volume is infused. “KVO: DOSE
COMPLETE” is shown on the Information Screen.
1) Press the
H
OLD
key
2) Press the
D
ATA
keys to put in a new volume
3) Press
R
UN
key to start infusion
Exit DOSE Mode:
1) Press
H
OLD
key
2) Press
M
ENU
key
3) Press
C
URSOR
keys to select "Exit DOSE Mode"
4) Press
E
NTER
key. The pump will return to the standard Hold state
DOSAGE CONC BODY
mcg/kg/min mg kg
change move
Press ENTER when done