Horizon Nxt®
Modular Infusion System 8 - 6 6/00
950520 Rev F
electrical outlet before cleaning.
Do not use the pump in the presence of strong electromagnetic fields. The
field may permanently damage the pump. For further information, consult
the manufacturer of the equipment in question.
To avoid mechanical or electronic damage, do not steam autoclave or
immerse the pump in any fluids or cleaning solutions.
A possible electrical shock hazard exists if the case of the pump is opened
or removed. Refer all servicing to authorized service personnel.
Do not use visibly damaged pumps or pumps with bent, damaged or missing
components.
Pumps should be tested for proper performance semiannually.
Federal (U.S.A.) law restricts this device to sale by or on the order of a
physician.