14
Aesculap
®
Full HD 3CCD Camera system PV460/PV462/PV463
7.2
Maintenance
To ensure reliable operation, the product must be maintained in accor-
dance with the maintenance label.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Note
It is recommended that the user should test the camera for operability and
also visually inspect the camera and its accessories (e.g., power cord,
cables) prior to each use as outlined in the section above.
Note
In your inquiries or correspondence please always indicate the model and
serial number printed on the identification plate. Further documentation is
available from the manufacturer on request.
If there is insulation damage to the power cord or accessory cables or if
any of the wall or equipment plugs are damaged, please replace the part
and dispose as outlined in this manual, see Disposal.
7.3
Safety checks
Safety checks require a functional test and electrical safety tests once a
year. It is not necessary that technical and safety inspections are per-
formed by qualified personnel; however, technical and safety inspections
of the camera and its accessories must be performed by persons, who,
based on their training, knowledge, and practical experience, are capable
of adequately performing such inspections and who do not require
instructions with regards to technical and safety inspection.
Required electrical safety according to IEC/DIN EN 60601-1 clause
18 & 19 or IEC 62353
CAUTION
No modification of this equipment is allowed! This
product has no user serviceable parts.
►
Do not modify this equipment without authori-
zation of the manufacturer.
►
If any part of the control unit and camera head
are damaged, contact your representative.
►
Do not remove the cover or attempt to do any
repairs yourself.
WARNING
Risk of injury and/or malfunctions!
►
Don’t use the device if technical and safety
inspection reveals a defect which could harm
the patient, clinicians, or third parties.
►
Don’t use the device until it has been properly
repaired
►
Immediately notify your representative of these
defects.
Tests
Requirements
Impedance between earth
connection and housing
≤
0.1
Ω
Enclosure leakage current
≤
0.1 mA
Earth leakage current
≤
0.5 mA
Patient leakage current
≤
0.1 mA
Patient auxiliary current D.C.
≤
0.01 mA
Patient auxiliary current A.C.
≤
0.1 mA