5
4.1
Connecting the accessories
Combinations of accessories that are not mentioned in the present
instructions for use may only be employed if they are specifically intended
for the respective application, and if they do not compromise the perfor-
mance and safety characteristics of the products.
All devices connected at the interfaces must demonstrably meet the
respective IEC standards (e.g., IEC 60950 for data processing equipment,
IEC/DIN EN 60601-1 for electromedical devices).
All configurations must meet the system standard IEC/DIN EN 60601-1-1.
The person connecting the units is responsible for the configuration and
must ensure compliance with system standard IEC/DIN EN 60601-1-1 or
equivalent national standards.
►
If you have any questions about that, contact your Aesculap partner or
Aesculap customer service.
4.2
Setting the correct voltage
The mains voltage must correspond to the voltage indicated on the unit
back panel.
4.3
Connecting the power supply
►
Plug in the power cord at the mains connection
25
of the camera con-
trol unit
10
.
►
Connect the mains plug of the power cord to the building mains.
►
Plug in the equipotentialization cable at the equipotentialization
connector
26
of the camera control unit.
►
Connect the plug of the equipotentialization cable to the building
mains equipotentialization connector.
►
To disconnect the camera from the mains, pull out the power cord from
the mains connection
25
.
4.4
Connecting the video cable
►
Connect the video cable at the camera back panel and the monitor or
the documentation system etc. Depending on the choice of the monitor
and documentation system, use the following camera outputs:
HD Monitor (High Definition):
■
DVI-D for digital HD 1080p quality (recommended)
– or -
■
HD-SDI for digital HD 1080i quality
– or -
■
S-Video for analog SD quality
5.
Working with the Full HD camera
5.1
System set-up
►
Position the camera as close as possible to the patient so that the cam-
era cable length of 4 m can be fully utilized.
►
Connect camera plug
9
to camera socket
17
at the front panel of cam-
era control unit
10
.
5.2
Function checks
►
Check camera control unit
10
and camera head
6
for external damage
and signs of knocks or other violent dents.
►
Check that camera cable
5
is not broken, kinked or twisted.
►
To switch on the camera system, press pushbutton
19
.
The current camera settings are displayed on the monitor:
AESCULAP Full HD 3 Chip Camera
MODE (current mode)
PERFORM WHITE BALANCE
►
Make certain the front window of the camera head
6
is absolutely
clean. If necessary, clean the window with isopropyl alcohol (70 %).
►
If applicable, check that the zoom adjusting ring turns without resis-
tance.
►
Make certain that focusing ring turns freely without sticking.
►
Align camera head
6
to an object in the room, focus the image and
check for good image quality.
►
To finish using the Full HD camera system, press pushbutton
19
.
WARNING
Risk of injury and/or product malfunctions due to
incorrect operation of the medical electric system!
►
Adhere to the instructions for use of any medi-
cal device.
WARNING
Risk of electric shock!
►
Only connect the equipment to a mains supply
with protective earth.
WARNING
Risk of injury and/or malfunctions!
►
Carry out function checks before every use.