5.13 Oscilloscope
TRIGGER EVENT:
Enabling trigger event it is possible to acquire and visualize the traces only at the occurrence of a
definite signal in Channel 1; that signal is characterized by a ring edge or a falling one and by a
level (or amplitude). To enabled a trigger event it is necessary:
1st to set the rising or falling edge (
Edge
icons);
2d to set the desired level (
Level
parameter).
Clicking the
Auto
button it is possible to disable the trigger event
�
the oscilloscope will continue
to acquire new data and update the traces.
You should use the
Auto
trigger function:
• During first acquisition, in order the know the scale of input signals;
• In presence of low repetitive signal rates;
• In presence of dc signals.
TIME BASE:
Time Base
This allows you to change the scale of the horizontal axis, the time base. The min. resolution is
1ms/div, while the max is 1s/div.
Delay
If the trigger event is enabled the value set in
Delay
fixes the point, in the horizontal axis, where
the trigger event will be visualized; on the contrary, if the trigger event is disabled the value set in
Delay is ignored.
The default value set for the Delay parameter is in the middle window.