![Avantes AVASPEC-1024 Скачать руководство пользователя страница 92](http://html1.mh-extra.com/html/avantes/avaspec-1024/avaspec-1024_manual_3025469092.webp)
Apr-09
91
Avantes
www.avantes.com
Power LED
When the 12V power supply is connected to the unit and the power switch is ON, the LED indicator is
lighted.
Rear Panel
Power Input Connector
Insert the 12VDC-power supply plug of the adapter into the unit Polarity on the plug is +inside, GND
outside.
Table of DB-15 connector
Pin
Description
10
GND
13
Shutter Open/Close
(High = Open, Low = Close)
Replacement of Lamps
The lamp unit consists of a deuterium lamp and a tungsten lamp, which are both installed in a shell
made of polytetrafluoroethylene (PTFE). If the deuterium lamp or the tungsten lamp fails or has
reached the end of its service life, the complete lamp unit has to be replaced.
If the lamp fails or has reached the end of its service life, the complete lamp unit can be easily
exchanged by the user and replaced by a new one.
For the replacement of the lamp unit (Replacement Lamp) the customer should follow the below
listed instructions:
1.
Disconnect power supply from the unit
2.
Open the housing by unscrewing 2 screws at front panel
3.
Take out the lamp unit
4.
Unscrew the allen screw (1.5 mm) in lamp box.
5.
Pull out the PTFE-housing upwards (do not tilt the housing)
6.
Remove the black plug from the tungsten lamp.
Replacement Lamp build-in
1.
Connect black plug onto tungsten lamp.
2.
Insert replacement lamp into the lamp box.
3.
Pay attention, that the contact pins on the bottom of the PTFE-body are
upright and do glide into the contact
sockets on the board.
4.
Tighten the allen screw again.
5.
Slide lamp back in outer box housing and
connect screws at front panel