Specifications / Parts List
Deep UV Deuterium
Deuterium Lamp
Halogen Lamp
Wavelength Range
190-400nm
215-400nm
360-1500nm
Warm-up Time
30 min.
30 min.
20 min.
Lamp Power
78W / 0.75A
78W / 0.75A
5W /0.2A
Lamp Lifetime
1000 hrs
1000 hrs
1000 hrs
Noise (AU)
2.10
-4
2.10
-4
10
-4
Max. drift
± 0.5%/hrs
± 0.5%/hrs
± 0.5%/hrs
Color Temperature
-
-
3000 K
Optical Power in
200μm fiber
11 µW
7 µW
43
µ
W
Optical Power in
600μm fiber
72 µW
61 µW
239 µW
Optical Power in
1000
μ
m fiber
206 µW
166 µW
354 µW
Power consumption
90 Watt (190Watt for heating D-Lamp 4-5 sec.)
Power Requirements
100-240VAC 50/60 Hz
Dimensions / Weight
315 x 165 x 140 mm / ca 5 kg.
Trouble Shooting
Deuterium Lamp Only
If the power supply or lamp does not seem to functioning properly, check the following:
FAULT
POSSIBLE CAUSE
REMEDY
Power switch on, fan not
working, nor one of the LED's
or lights
Line power not present
Check line voltage
Fuse(s) defective
Check /replace fuse(s)
Deuterium Lamp does not
ignite.
The Two-Color-LED of
deuterium lights up red
(ERROR)
Deuterium Lamp too hot
Allow Deuterium Lamp to cool
down.
Swich to OFF to reset the
Deuterium lamp, then switch to
ON to restart.
Lifetime of Deuterium
Lamp has expired
Replace Deuterium Lamp
Deuterium Lamp internal
connection plug is not
connected right
Disconnect power, Open Unit and
connect connector plug
Deuterium Lamp extinguishes
When this fault appears always switch off the deuterium power