76
Apr-09
Avantes
www.avantes.com
View Menu: Auto scale Y-axis
By using this option, the graph will be rescaled on-line. A maximum signal will be shown at about
75% of the vertical scale. This will be realized for all channels in the current view mode.
View Menu: Goto Preset Scale
By clicking this menu option, the scale for X- and Y-axis will be set to a range that has been set
before . The same result can be achieved by clicking the Goto Preset Scale Button in the button bar.
View Menu: Grid Enable
With the Grid Enable option activated, a grid will be displayed in the graph as shown in the figure
below.
View Menu: Progress Bar Enable
If using long integration times or a high number of averages, it can take a few or more seconds
before a new scan is received by the application. To get an indication about how much time it will
take until the next scan arrives, a progress bar can be displayed. After enabling the progress bar by
clicking the menu option, it will be displayed after the next scan has arrived. The progress bar will
be shown only if the time between scans is more than one second. The time between scans is
roughly the integration time, multiplied with the number of averages. However, if the number of
averages is high, the time between scans can get longer because of the overhead time that is spent
on transmitting the high number of average spectra to the PC.