Relationship between trigger conditions
A relationship between trigger conditions on edges and trigger conditions on logic levels can be set
by the user, by using the
Trigger condition
menu ring. Several options are available:
Edge OR Level
A trigger condition occurs when one between an edge
condition or a logic level condition is detected
Edge AND Level
A trigger condition occurs when an edge condition and a logic
level condition are satisfied simultaneously
Edge
BEFORE
Level
A trigger condition occurs when an edge condition has been
detected at least once before that on logic levels
Edge AFTER Level
A trigger condition occurs when an edge condition has been
detected at least once after that on logic levels
ALWAYS
A trigger condition occurs automatically when data acquisition
starts. Any setting on edges or on logic levels will be ignored
NEVER
Trigger is inhibited. Any setting on edges or on logic levels
will be ignored. A trigger event, however, can be activated
manually during data acquisition