When more than one data condition are set, the
Conditions Operator
control is enabled, by which is
possible to choose between:
•
Sequence
: the AT-LA500 triggers when the address condition and all the entered data
conditions are verified in the specified order;
•
OR
: the AT-LA500 triggers when the address condition and at least one of the entered data
conditions are verified.
Pretrigger
The
Pretrigger
control on the bottom side allows specifying the position in which the trigger event
will be located with respect to the AT-LA500 memory. If the most meaningful data must be
sampled after the trigger detection, low pretrigger percentage must be selected. On the contrary, if
signal evolutions before trigger event must be analyzed, a high pretrigger percentage is more
suitable.