AT-LA500
–
U
SER MANUAL
R
ELEASE
3.4
–
JUNE
2009
32
I
2
C Trigger
In this mode, the AT-LA500 can trigger when a selected event occurs on a I
2
C (Inter Integrated
Circuit) bus. Furthermore, the acquired data are decoded and shown in the Waveform View and
State Listing windows.
To set up the AT-LA500 for an I
2
C acquisition, the following steps are required:
•
Connect a Passive Probe or a High-Z Probe to the pod A. In this case you also have to
correctly set the threshold voltage in the
Probes
tab of the
Settings
window;
•
Connect at least one grounded pin (white wires) to the ground of the board to test;
•
Connect the
Channel 0
(black wire) to the
SDA
line of the I
2
C bus to test;
•
Connect the
Channel 1
(brown wire) to the
SCL
line of the I
2
C bus to test.
Note
: It is not possible to use a Low-C Probe to analyze an I
2
C bus: this could prevent the system
under test to work properly.
The I
2
C trigger has 2 different options: Address and Address&Data. These options are selectable by
using the
I2C Trigger Type
control.
Address
In this mode, the AT-LA500 can trigger when the I
2
C bus addresses one or more peripheral.
In the
Trigger On
control, it is possible to choose which operation has to be triggered. The available
options are:
•
Read or Write
: trigger on every operation on the bus;
•
Read
: trigger when the bus is performing a read operation, ignoring the ACK bit;
•
Write
: trigger when the bus is performing a write operation, ignoring the ACK bit;
•
ACK R/W
: trigger on every acknowledged operation on the bus;
•
ACK Read
: trigger on an acknowledged read operation on the bus;
•
ACK Write
: trigger on an acknowledged write operation on the bus;
•
NACK R/W
: trigger on every not acknowledged operation on the bus;
•
NACK Read
: trigger on a not acknowledged read operation on the bus;
•
NACK Write
: trigger on a not acknowledged write operation on the bus;