MAGB2 User Guide
36
V1.7 2021-10-01
Delete Total + Volume
–
write value different to zero for erasing the Total + flow totalizer.
Delete Total Volume
–
write value different to zero for erasing the Total flow totalizer.
Flow Simulation
– switch off/on the simulation flow function. Default OFF.
Simulated Flow
– write simulated flow.
Service Mode
– write value different to zero for turn ON the service mode. Default OFF.
Measurement
– 0 = Stop – the unit shows actual flow, but the totalizers are stopped.
1 = Running
– totalizers are active. Default Stop.
Air Constant
– constant value to determine the Empty pipe detection limit. Default 188.
Invert Flow
– this function serves to change the direction of the flow. Default OFF.
Low Flow Cutoff
– this function serves to set the minimum flow the flowmeter will react on. Default 2%.
Flow Qn
– setup to the excepted flow Qn. It is set automatically when you write diameter.
Date Settings
- date write in BCD format YYYYMMDD (ie 25.03.2010 = (hex)0x20100325)
Time Settings
- time write in BCD format HHMMSS
(ie 08:33:15 = (hex)0x00083315)
Battery insert date
– each time main battery is replaced it is necessary to set correct date of replacement
Battery capacity left
– each time main battery is replaced it is necessary to set correct battery capacity
Delete OK (min)
–
write value different to zero for erasing the OK min counter.
Delete Error (min)
–
write value different to zero for erasing the Error min counter.