MAGB2 User Guide
30
V1.7 2021-10-01
11.8. Display
Modbus
register
Modbus
address
Data
type
No. of
decimal
Min.
Value
Max.
Value
Default Description
Read/
Write
1500
1499
Longint
-
0
4
2
Unit Flow (+/-),
0=UKG/min,
1=USG/min, 2=m3/h,
3=l/min,
4=l/s
R/W
1502
1501
Longint
-
0
3
2
Unit Volume, 0=UKG,
1=USG,
2=m3,
3=l
R/W
1504
1503
Longint
-
0
1
0
Unit Temperature
0 = °C,
1 = °F
R/W
1506
1505
Longint
-
0
1
0
Unit Pressure
0 = bar,
1 = psi
R/W
1508
1507
Longint
-
0
5
0
Language*,
0 = ENG
R/W
1510
1509
Longint
-
0
100
25
Contrast [%]
R/W
1512
1511
Longint
-
0
30
0
Backlight
R/W
1514
1513
Longint
-
0
2147480
30
Display Enable Time
[s]
R/W
*so far only ENG implemented, other languages will come in future updates
Unit Flow
– actual flow unit (default m
3
/h)
Unit Volume
– totalizer unit (default m
3
)
Unit Temperature
– applies for external temperature sensor (default °C)
Unit Pressure
– applies external pressure sensor (default bar)
Language
–
Language of transmitter menu. (so far English only)
Contrast
– Set display’s contrast
Backlight
–
sets the time in seconds when the backlight is on. 0 means backlight off.
Display Enable Time
–
number of seconds when the display is working after touching the button. This
function is for battery conservation. Value must be multiple of 5s. (If you write bad value, non multiple, this
value is automatically converted). If the value is “0”, function is off and display is all the time
ON
.