MAGB2 User Guide
35
V1.7 2021-10-01
11.11. Service settings
To enter this section, it is necessary to enter the Service Password.
Modbus
register
Modbus
address
Data
type
No. of
decimal
Min
Value
Max Value
Default
Description
Read/
Write
3000
2999
Longint
0
0
9999999
120000
Noise filter
threshold
R/W
3002
3001
Bool
0
0
1
0
Start noise
filter
R/W
3004
3003
Bool
0
0
1
0
Delete Error
R/W
3006
3005
Bool
0
0
1
0
Delete Total - R/W
3008
3007
Bool
0
0
1
0
Delete Total
+
R/W
3010
3009
Bool
0
0
1
0
Delete Total
R/W
3012
3011
Bool
0
0
1
0
Flow
Simulation,
0=OFF,
1=ON
R/W
3014
3013
Longint
3
0
36 000 000
3600
Simulated
Flow
R/W
3016
3015
Bool
-
0
1
0
Service
Mode,
0=OFF,
1=ON
R/W
3018
3017
Bool
0
0
1
0
Measuremen
t,
0=Stop,
1=Running
R/W
3020
3019
Longint
0
-
9999999
9
99999999
450000
0
Air Constant
R/W
3022
3021
Bool
0
0
1
0
Invert Flow,
0=No-invert,
1=Invert
R/W
3024
3023
Longint
0
0
5
3
Low Flow
Cutoff,
0=OFF,
1=0.5%,
2=1%,
3=2%,
4=5%,
5=10%
R/W
3026
3025
Longint
3
0
36 000 000
-
Flow Qn
R/W
3028
3027
Longint
0
0
0x2999123
1
-
Date Settings R/W
3030
3029
Longint
0
0
0x0023595
9
-
Time
Settings
R/W
3032
3031
Longint
0
0
0x2999123
1
-
Battery insert
date
R/W
3034
3033
Longint
0
0
999999
-
Battery
capacity left
R/W
3036
3035
Bool
0
0
1
0
Delete Error
(min)
R/W
3038
3037
Bool
0
0
1
0
Delete Ok
(min)
R/W
Noise filter threshold
– determines which value from AD converter is restarting the averaging filter
Start noise filter
– manual restart of the filter
Delete Error
– major errors needs to be deleted by service settings
Delete Total
– Volume –
write value different to zero for erasing the Total
– flow totalizer.