MAGB2 User Guide
29
V1.7 2021-10-01
11.7. Info
Modbus
register
Modbus
address
Data type
No. of
decimal
Min
Value
Max Value
Description
Read/
Write
1000
999
Longint
0
0
0x29991231
Date
R
1002
1001
Longint
0
0
0x00235959
Time
R
1004
1003
Longint
0
0
999 999
Unit No.
R
1006
1005
Longint
0
0
2^32
Error code R
1008
1007
Longint
0
0
1 000
Diameter
R
1010
1009
Longint
3
0
36 000 000
Flow Qn
R
1012
1011
Longint
2
0
9 999
FirmWare
No.
R
1014
1013
Longint
0
0
2^32
Excitation
counter
R
1016
1015
Longint
0
0
65535
FW
Checksum
R
1018
1017
Longint
0
0
0x29991231
Battery
Insert Date
R
1020
1019
Longint
0
0
2^32
Battery
Remainig
Capacity
R
1022
1021
Longint
0
0
100
Battery
Status
R
1024
1023
Longint
0
0
2^32
Error (min) R
1026
1025
Longint
0
0
2^32
OK (min)
R
Date
- date is stored in BCD format YYYYMMDD
(ie 25.03.2010 = 0x20100325)
Time
- time is stored in BCD format HHMMSS (ie 08:33:15 = 0x00083315)
Unit no.
– exclusive number for this Flowmeter. If there are any problems, please refer to this number.
Error code
– Convert read value to binary number. Number one means error. For more information see
chapter 19.
Diameter
– this item shows the nominal sensor diameter that is currently configured for the given
flowmeter.
Flow Qn
– nominal flow of the meter, set around 3m/s.
Firmware No
.
– this shows the current firmware version.
Excitation Counter
– the number of excitations after battery exchange.
FW Checksum -
validation of not allowed changes in FW (metrological purposes).
Battery Insert Date
– date of the last battery exchange. in BCD format YYYYMMDD (ie 25.03.2020 =
0x20200325)
Battery Remaining Capacity
– remaining battery capacity in mAh.
Battery Status
- this number shows remaining capacity of battery in %.
Error
(min)
– the number of minutes the device was not measuring because of errors.
OK
(min) - the number of minutes that the device measured correctly.