MAGB2 User Guide
14
V1.7 2021-10-01
5. Frequency output
As an analogue output MAGB2 has one frequency output (FOUT) configurable to be flow dependent. The
output is galvanically isolated open drain with following connection. The use is mainly for calibration
purposes, can be used as flowrate indicator for PLC units.
Connection:
Recommended resistance R=1 k Ω
The output can be set to:
0) OFF
(default)
1) FREQUENCY OUT
– the output generate frequency based on actual flow. User sets the flow at 1000Hz
signal. Then the frequency is proportional to the flow. The battery current consumption is higher if you
choose this option.
2) PULSE VOLUME
–
this setting works the same way as relay output->volume pulse: the output
generates a pulse when the preset volume has passed the sensor (the pulse width can be set). The battery
current consumption is higher if you choose this option. Minimal resolution is 1ml.
6. Relay output
As a second analogue output MAGB2 has one latching relay output configurable to an error detection or
flow dependent. This output is used for status or volumetric information. The output is galvanically isolated
dry contact.
Connection:
Frequency
output
connector
RELAY
output
connector