60
Citadel Bed Frame System
Symptom
Possible Cause
Action
Weighing system display
shows unknown value
System fault or operator
error
Refer to the table of error codes on
page 41
TV / lighting controls not
working
Nurse call cable not
connected or wrong type
Check to ensure cable is of correct
type and properly connected
Error Codes
Display
Error Name
Description
Action
E001
Zeroing the bed
on battery
Zero on battery
power
Connect bed to mains supply and
re-initalize the weighing system
E002
Unsteady weight
Error when auto
compensation weight
is lower than minus
100 kg (220 lb)
The E002 error will be cleared
by adding weight until the auto
compensation is above -100 kg
and then pressing the auto-comp
key (H1) again, or when weight is
removed and weighing zeroed.
E003
Error when auto
compensation
weight is higher
than minus 100
kg (220 lb)
The auto
compensation weight
of 100 kg (220 lb)
change has been
exceeded and the
auto compensation
will not be completed
Error will be cleared by removing
weight until the auto compensation
is below 100 kg (220 lb)
E102
Weigh time-out
(10 seconds)
During a weight
command, the weight
was not stable and
changing too much
The bed was touched or leaned
on, repeat with no one touching
the bed
E300
Button stuck error
Control button
depressed for more
than 90 seconds
Remove pressure from control
buttons
E410
General Error
Service error which requires technical investigation
Содержание citadel
Страница 1: ...Citadel Bed Frame System Instructions for Use with people in mind 830 331 Rev A 08 2016 ...
Страница 72: ...72 Citadel Bed Frame System ...
Страница 73: ...73 Citadel Bed Frame System ...
Страница 74: ...74 Citadel Bed Frame System ...
Страница 75: ...75 Citadel Bed Frame System ...