39
Citadel Bed Frame System
Weighing Initialization
The weighing system MUST be zeroed every time a new patient
is to be placed on the bed.
The weighing system MUST be zeroed every time a mattress is
changed.
The weighing system cannot be zeroed if a mattress or
accessories weighing more than 50 kg (110 lb) are fitted.
It is not recommended to zero the bed or take a weight reading
in the lowest position.
After connecting the bed to the electricity supply, the weighing system must be
initialized as follows:
1. Place the mattress, bed linen and all necessary accessories on the
bed. The patient must not be on the bed at this point.
2. Press the Zero button once.
3. After a few seconds the display will read zero .
4. The patient can now be placed on bed.
5. The bed cannot be re-zeroed once a patient is on the bed. Use the Auto
Compensation feature if adding or removing items from the bed such as
linen or accessories
Weighing
To calculate and display the patient’s weight:
1. With the patient centered on the bed, press the Weigh
button once.
2. The display will show a moving circular pattern until a
stable weight reading is obtained.
3. The patient’s weight will be displayed for ten seconds then
the display will go blank.
Scale Readings - Scales / patient weights are for reference only.
Scale readings should not be relied upon for medication dosage.
All equipment on the weighed portion of the unit is included in
weight displayed.