51
Citadel Bed Frame System
NURSING CARE
It is recommended that all chapters of this manual be reviewed prior to product
use. Carefully read the
Contraindications
,
Risks and Precautions
and
Safety Information
sections of the
Introduction
chapter of this manual prior to
performing nursing care for a patient on the
Citadel
Bed Frame System.
CPR
The following steps indicate how to place the bed into the CPR position.
1.
Press and hold the CPR button.
This will immediately flatten
the head and knee sections and return the frame to level from
the Trendelenburg or reverse Trendelenburg positions. If the
bed is set to a height > 480mm (19 in), then activating the CPR
function will also lower the bed to 480mm (19 in). The CPR
buttons will flash on all control panels and an audible tone will
sound. If installed, the
Citadel
Patient Therapy System will
deflate the mattress and turn off.
2. Lower side rails.
3. Remove headboard, if necessary.
4. Depending on the patient support surface in use, a crashboard
may be needed.
5. Begin CPR. Follow institution CPR guidelines.
6. Reinstall headboard, if necessary.
7. Raise and lock side rails.
8. If applicable, resume therapy as directed by physician.
9. Adjust patient surface for patient comfort.
In a fault or power loss condition where the CPR button is not responding, use
the CPR backrest release to position the patient for CPR. If installed, the
Citadel
Patient Therapy System will deflate the mattress and turn off.