11.2.3 Connect the four actuator cables (“Electrical Assembly” on page 22, Item 15/16/17/18) and
OpenBus cable (“Electrical Assembly” on page 22, Item 19 and 20) to the control box, making
sure each cable is connected to the correct socket as previously noted. Thread the retainer
(Fig 10, Item A) over the cables and clip it into place.
11.2.4 Reconnect the backup battery (Fig 6, Item 4) to the control box, refer to Subsection 9.2.
11.2.5 Reconnect the mains lead (“Electrical Assembly” on page 22, Item 9) to the control box, refer
to Subsection 10.2.
11.2.6 Using the electrical safety test equipment (see “Tools and Equipment” on page 21), verify that
the earth bonding resistance value is within the limits shown in “Technical Data” on page 91.
Содержание Enterprise E8000X
Страница 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Страница 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Страница 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Страница 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Страница 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Страница 33: ...Enterprise E8000X Service Manual Servicing Instructions Page 29 Figure 3 Sub Frame Sections Assembly...
Страница 62: ...Servicing Instructions Enterprise E8000X Service Manual Page 58 This Page Intentionally Left Blank...
Страница 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Страница 87: ...Enterprise E8000X Service Manual Servicing Instructions Page 83 Figure 28 Location of calf section stay A B...
Страница 104: ...This Page Intentionally Left Blank...