Guidance and manufacturer’s declaration – electromagnetic immunity
The bed is intended for use in the electromagnetic environment specified below. The customer or the
user of the bed should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 610000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power
supply lines
± 2kV
Mains power quality should be that of a
typical commercial or hospital
environment.
± 1kV for input/
output lines
Not applicable
Surge
IEC 61000-4-5
± 1kV line to
line
± 1kV
Mains power quality should be that of a
typical commercial or hospital
environment.
± 2kV line to
earth
± 2kV
Voltage dips,
short interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
70%
U
T
(30% dip in
U
T
)
for 25 cycles
40%
U
T
(60% dip in
U
T
)
for 5 cycles
<5%
U
T
(>95% dip in
U
T
) for 0.5 cycle
<5%
U
T
(>95% dip in
U
T
) for 5 sec.
70% of input
voltage for 500ms
40% of input
voltage for 100ms
<5% of input
voltage for 10ms
<5% of input
voltage for
5000ms
Mains power quality should be that of a
typical commercial or hospital
environment.
If the user of the bed requires continued
operation during power mains
interruptions, it is recommended that the
bed be powered from an uninterruptible
power supply or battery.
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3A/m
3A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE
U
T
is the a.c. mains voltage prior to the application of the test level
Содержание Enterprise E8000X
Страница 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Страница 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Страница 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Страница 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Страница 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Страница 33: ...Enterprise E8000X Service Manual Servicing Instructions Page 29 Figure 3 Sub Frame Sections Assembly...
Страница 62: ...Servicing Instructions Enterprise E8000X Service Manual Page 58 This Page Intentionally Left Blank...
Страница 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Страница 87: ...Enterprise E8000X Service Manual Servicing Instructions Page 83 Figure 28 Location of calf section stay A B...
Страница 104: ...This Page Intentionally Left Blank...