Enterprise E8000X Service Manual
Troubleshooting
Page 13
3
Fatal Error Troubleshooting
3.1 Installation
The hardware is available under index S9411 and contains (SDT box and RJ45 10-wires cable).
Standard cable USB A-B is not supported.
NOTE
Do not connect SDT together with Linak programming tool
1.
Download installation file. For downloading contact product technical support group
(Arjo manufacturing site).
2.
Add CB60BM extension file by choosing Configuration -> Import.
3.
SDT has built-in help, to get information about any function press F1 in selected sheet.
3.2 Diagnostics
1.
Connect SDT to OpenBus to diagnose the bed. Use free port of MJB or HB port of CB6.
2.
Connect SDT to PC, open program and choose CB60BM Control Box.
3.
Startup check for verifying the fundamentals of the system:
• Check if OpenBus is running. Open SDT and check Connection Info and Devices.
• Check measured Info if 8 V and 40 V light green. It is not OK if it is blinking and it
could cause the following: feedback error, false Endstop detection. Indicators blinking
on Control Panels, communications missing.
• Check if Power REQ is activated only when a button on Control Panels is activated.
3.3 Statistics
Program allows the verification of statistics data of OpenBus system.
3.4 Channel State
3.4 Overall Status
Activations In
Number of actuators activations inward direction
Activations Out
Numbers of actuators activations outward direction
Endstop In
Number of Endstop In registered during use
Endstop Out
Number of Endstop Out registered during use
Overloads
Number of actuators overload
Accumulated Work [A*s)
Shows accumulated work (current consumed * time)
Accumulated duration [s]
Shows accumulated duration of activity inwards and outwards
direction
Errors
Number of actuator errors
Power request errors
Communication error CB - OpenBus component
FET errors
Number of actuator H-Bridge errors
Flash Write Counter
Number of Flash program
Содержание Enterprise E8000X
Страница 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Страница 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Страница 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Страница 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Страница 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Страница 33: ...Enterprise E8000X Service Manual Servicing Instructions Page 29 Figure 3 Sub Frame Sections Assembly...
Страница 62: ...Servicing Instructions Enterprise E8000X Service Manual Page 58 This Page Intentionally Left Blank...
Страница 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Страница 87: ...Enterprise E8000X Service Manual Servicing Instructions Page 83 Figure 28 Location of calf section stay A B...
Страница 104: ...This Page Intentionally Left Blank...