17
Thigh Actuator - Replacement
This section describes how to replace the thigh actuator. Reference is made throughout this section to
“Electrical Assembly” on page 22 and Figure 3 on page 29.
17.1
Removal
To remove the Thigh Actuator:
17.1.1 Raise the bed to maximum height to allow easy access to the underside of the mattress
platform. Release the cable clip from the thigh actuator (Fig 3, Item 8) and unplug the actuator
cable (“Electrical Assembly” on page 22, Item 15) refer to Subsection 1.2.
17.1.2 Pry off and discard the starlocks (Fig 3, Item 10) from both ends of the thigh actuator. Attach
the hoist to the foot end of the mattress platform and slowly raise the bed until the thigh actuator
just becomes free on its mountings.
17.1.3 Knock out two clevis pins (Fig 3, Item 6) to release the actuator from the thigh section
(Fig 3, Item 12) and sub-frame assembly (Fig 3, Item 11). Collect two shakeproof washers
(Fig 3, Item 9).
17.2
Installation
To install the Thigh Actuator:
17.2.1 Position the thigh actuator (Fig 3, Item 8) on the bed with the ram pointing towards the foot
end. Attach the actuator body to the sub-frame assembly (Fig 3, Item 11) with the clevis pin
(Fig 3, Item 6) and shakeproof washer (Fig 3, Item 9). Secure the assembly with a new
starlock (Fig 3, Item 10). Make sure the shakeproof washer and starlock fit tightly against the
mounting bracket – see Figure 16.
17.2.2 Position the thigh section (Fig 3, Item 12) to align the mounting lug with the hole in the thigh
actuator ram. Attach the actuator ram to the thigh section using the clevis pin (Fig 3, Item 6)
and shakeproof washer (Fig 3, Item 9) and secure with a new starlock (Fig 3, Item 10). Make
sure the shakeproof washer and starlock fit tightly against the mounting bracket – see
Figure 16.
17.2.3 Connect the thigh actuator cable (“Electrical Assembly” on page 22, Item 15). Make sure the
connector is fully seated in the socket and refit the cable clip, refer to Subsection 1.2.
WARNING
Ensure that the foot end of the mattress platform is securely supported before
removing the thigh actuator.
Содержание Enterprise E8000X
Страница 1: ...746 587_5 09 2021 SERVICE MANUAL Enterprise E8000X E8X For internal use only...
Страница 8: ...Introduction Enterprise E8000X Service Manual Page 4 This Page Intentionally Left Blank...
Страница 10: ...Decontamination Enterprise E8000X Service Manual Page 6 This Page Intentionally Left Blank...
Страница 18: ...Troubleshooting Enterprise E8000X Service Manual Page 14 3 5 Fatal Error H Bridge Diagram...
Страница 19: ...Enterprise E8000X Service Manual Troubleshooting Page 15 3 6 Fatal Error Actuator Diagram...
Страница 33: ...Enterprise E8000X Service Manual Servicing Instructions Page 29 Figure 3 Sub Frame Sections Assembly...
Страница 62: ...Servicing Instructions Enterprise E8000X Service Manual Page 58 This Page Intentionally Left Blank...
Страница 69: ...Enterprise E8000X Service Manual Servicing Instructions Page 65...
Страница 87: ...Enterprise E8000X Service Manual Servicing Instructions Page 83 Figure 28 Location of calf section stay A B...
Страница 104: ...This Page Intentionally Left Blank...