If the result of any of these tests is unsatisfactory, contact Arjo
or an approved service agent.
Weighing system
Be sure to follow all local guidelines or regulations regarding weigh
scale verification and calibration.
The weigh system should be re-verified before the date of expiry
shown on the bed. Make sure the bed is re-verified every 12
months. If verification does not pass, the weigh system should be
re-calibrated.
Refer to the product service manual for verification and calibration
procedures or contact an approved Arjo service agent.
WARNING
The procedures below must be carried out by suitably trained and
qualified personnel. Failure to do so may result in injury or an unsafe
product.
Actions to be done by qualified personnel
Yearly
Check that the bedstripper (linen shelf) remains in its closed
position when maximum foot down tilt is applied
9
Check that the bed operates correctly using the backup battery
as described in the section “Battery test” on page 61.
9
Check operation the of castors, paying special attention to
braking and steering functions
9
Check that the bed extension locks securely in all three positions
9
Examine the power supply cord and mains plug; if damaged,
replace the complete assembly; do not use a rewireable plug
9
Examine all accessible flexible cables for damage and
deterioration
9
Check all accessible nuts, bolts and other fasteners are present
and correctly tightened
9
Check any accessories fitted to the bed, paying particular
attention to fasteners and moving parts
9