51
SafeSet™
SafeSet
visual status indicators provide a quick visual
indication of optimum bed settings for patients identified as
being at risk of falls.
SafeSet
visual status indicators show four safety-critical bed
parameters:
• Brake setting
• Side rail position
• Mattress platform height
• Movement detection status.
Two identical
SafeSet
indicator panels are located just below
the foot board.
Indicator lights (
1
) above each symbol light red to show a
potentially unsafe condition, or green to indicate a safe
condition.
Fig. 31 -
SafeSet
indicator panel
* To comply with FDA guidelines, the system will still show a “safe” condition if one (but not both) of the
foot end side rails are down.
WARNING
SafeSet
visual status indicators are only for use with patients at risk
of falls and should not be used with patients that are at risk of
climbing over the side rails.
Symbol
Indicator light green
(“safe” condition)
Indicator light red
(“unsafe” condition)
Brakes on
Brakes off
All side rails raised*
If either of the head end panels are
lowered.
If both foot end panels are lowered.
Mattress platform at
minimum height (or within
25mm of minimum height)
Mattress platform not at minimum height
VariZone patient movement
detection set
VariZone patient movement detection not
set
1
2