44
Accuracy
The weighing system is very sensitive and can be affected by
a number of external factors. For best performance, observe
the following precautions:
• Ensure the bed is on a flat, level surface.
• Ensure the transport locking bolts and washers have
been removed before use – refer to page 12.
• Position the bed so that it is clear of obstructions, e.g.
walls, furniture, cables and curtains.
• Ensure the patient is fully on the mattress during
weighing.
• While weighing is taking place, the patient should keep
as still as possible and other persons should keep clear
of the bed.
• Ensure that the patient, pillows and bed linen do not
touch the head board during weighing.
Fig. 28 - Areas of patient weighing influence
Caution
Any item attached to or placed on any part of the bed other than the
head board (Fig 28, Item 1) and head end lifting pole accessory sockets
(Fig 28, Item 2)
will influence the patient weighing system.
Make sure the mains cable is free of the head board and any fixed part
of the bed. If the cable becomes trapped it can affect the accuracy of
the weighing system.
1
2
Areas that do influence patient weighing
Areas that do not influence patient weighing
NOTE
Use the AUTO COMPENSATION feature before adding or removing any items (eg:
urine bags, iv poles, mattress pumps, bed-clothes etc.).