63
Error code E410 and
flashing control panel
lights
Flat battery
If the bed has just been
delivered or has been
left in storage for a long
period of time without
charge, plug the bed
into mains power and
leave to charge as per
instructions.
If the problem persists
call an Arjo approved
service engineer
Alarm sounds when
patient moves on the bed
Movement detection
threshold set too low
Increase VariZone patient
movement detection
threshold setting
Errors in indicated patient
weight
Mattress platform
obstructed
Check the mattress
platform is not touching
furniture, curtains, cables,
etc.
Check that weight has not
been added to the bed
without using the auto-
compensation function.
Nurse call function not
working
Nurse call cable not
connected or wrong type
Check cable is of correct
type and properly
connected
Backrest does not lower
using the manual CPR
release handle
Insufficient patient weight
applied to the backrest
Push the backrest down to
initiate lowering
Error code E001
Zero on battery power
Connect bed to mains
supply and re-initialise
the weighing system -
see page 45
Error code E002
Auto Compensation
weight decrease more
than 100kg (220 lb.)
Replace weight removed
from bed
Error code E003
Auto Compensation
weight increase more than
100kg (220 lb.)
Remove added weight
from bed
Error code E102
A stable weight reading
was not obtained within 10
seconds
Refer to the points listed
in the section
“Accuracy” on page 44
Symptom
Possible Cause
Action