64
Product lifetime
The lifetime of this equipment is typically ten (10) years. “Lifetime”
is defined as the period during which the product will maintain the
specified performance and safety, provided it has been
maintained and operated in conditions of normal use in
accordance with the requirements in these instructions.
TV/lighting controls not
working
Nurse call cable not
connected or wrong type
Check cable is of correct
type and properly
connected
Red brake light on
SafeSet
indicator panel
but the brakes are in the
on position
Brake sensor failure
Replace brake sensor
Red side rail light on
SafeSet
indicator panel
but the side rails are in the
raised position
Side rail sensor failure
Replace side rail sensor
No lights on
SafeSet
indicator panel
Bed disconnected from
electricity supply
Bed running in power
conservation mode.
Reconnect electricity
supply and swipe fob to
reactivate
SafeSet
indicator panel.
SafeSet
indicator panel
lights always bright
Light sensor obscured or
failed.
Remove any items
obscuring the
SafeSet
indicator panel light
sensor. If no improvement
call an Arjo approved
service engineer
Symptom
Possible Cause
Action