SECTION 7: PROCESSING SAMPLES FOR COLIFORM ANALYSIS
All samples must be incubated within 6 hours of sampling.
Dispense a sterile absorbent pad into a sterile petri dish, and saturate the pad with prepared
broth
Loosen the graduated filter funnel, and remove from the base support.
Sterilise the forceps and allow to cool. Using these forceps, place a sterile membrane onto
the glass membrane support, grid side up. If the membrane tears or becomes contaminated,
discard it and use a new one.
Lock the membrane in place by screwing the filter funnel down into position.
Pour the water sample into the filter funnel up to the 100 ml graduation.