2.3.7
Performing Tests
When zero calibration is complete, remove the vial from the sample chamber and perform the tests
as described in the relevant application note.
When the results have been displayed, options include:
-
Storing and/or printing the results
-
For some methods, changing between different chemical species
-
Performing further analysis with the same zero
-
Selecting a new method
2.3.8
Changing Chemical Species
For some methods, there it is possible to change the chemical species of the test results. When the
test result is displayed, press the arrow key [ ] or [ ].
Example:
320 Phosphate LR T
---- [ ] ----> 320 Phosphate LR T <---- [ ] ---- 320 Phosphate LR T
0.05-4 mg/l PO
4
0.02-1.3 mg/l P
0.04-3 mg/l P
2
O
5
1.00 mg/l PO
4
---- [ ] ---->
0.33 mg/l P
<---- [ ] ----
0.75 mg/l P
2
O
5
If the species of a test result is changed, the displayed range is adjusted automatically. For an already
stored result, it is not possible to change the chemical species. The last displayed chemical species is
kept by the instrument and will be displayed if this method is used again. If there is the possibility to
change the chemical species for a method, it is described in the application note. The arrows
indicate the possible chemical species and are printed below the application note.
PO
4
P
P
2
O
5
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...