Delete Stored Results
Press [MODE], [Shift] + [3] [4] keys.
Confirm with [ ] key.
The display shows:
Select the required method from the displayed list or enter the method
number directly.
Press [Shift] + [0] keys to retain the data sets in memory.
Press [Shift] + [1] key to delete all the data sets in memory
The display shows:
Press [ ] key to delete.
Press [ESC] to cancel without deleting data.
Note:
All stored test results are deleted.
2.4.4
Calibration
Calcium Hardness Method 191 – Calibration of a Method Blank
Press [MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [1] keys.
The display shows:
1. Fill a clean vial (24 mm Ø) with
exactly 10 ml of deionised water
, close
tightly with the cap.
2. Place the vial in the sample chamber, making sure that the triangular
marks on the vial and instrument are aligned.
3. Press [ZERO] key.
< Delete data >
Delete all data?
YES: 1
NO: 0
< Delete data >
Delete data
Do not delete: ESC
< Calibration >
1: M191 Ca-hardness 2
2: M191 reset 0 cali.
3: M170 Fluoride L
< Calibration >
191 Ca-hardness 2 T
prepare Zero
press ZERO
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...