Enter data of the coefficient A including decimal point
e.g. [Shift] + [1] [.][3][2]
Press [F1] key to reset numerical input.
Confirm with [ ] key.
The display shows:
Press arrow key [ ] or [ ] to change between plus and minus sign.
Enter the exponent of the coefficient A
e.g. [Shift] + [3]
Confirm with [ ] key.
Successively, the instrument queries the data for the other coefficients (B, C,
D, E and F).
Note:
If zero [0] is entered for the value of the coefficient, the input of the
exponent is omitted automatically.
Confirm every input with [ ] key.
The display shows:
Enter measurement ranges from -2600 mAbs to +2600 mAbs.
Press arrow key [ ] or [ ] to change between plus and minus sign.
Enter the values in Absorbance (mAbs) for the upper limit (Max) and the
lower limit (Min).
Confirm with [ ] key.
The display shows:
Press arrow key [ ] or [ ] to change between plus and
minus sign.
Confirm with [ ] key.
<User polynoms.>
A: 1.32___ E+____
<User polynoms.>
y = A+Bx+Cx2+Dx3+
Ex4+Fx5
B: + __________
<User polynoms.>
measurement range
Min mAbs: +_________
Max mAbs: +_________
<User polynoms.>
choose unit:
>>
mg/l
g/l
mmol/l
mAbs
µg/l
E
A
%
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...