Press [Shift] + [2] keys.
The display shows:
Press [Shift] + [0] keys to keep the method blank.
Press [Shift] + [1] keys to erase the method blank and set the value back to
factory calibration.
Fluoride Method 170
Press [MODE], [Shift] + [4] [0] keys.
Confirm with [ ] key.
The display shows:
Press [Shift] + [3] keys.
The display shows:
1. Fill a clean vial (24 mm Ø) with
exactly 10 ml of deionised water
, close
tightly with cap.
2. Place the vial in the sample chamber, making sure that the triangular
marks on the vial and instrument are aligned.
3. Press [ZERO] key.
The display shows:
4. Remove the vial from the sample chamber.
5. Add
exactly 2 ml SPADNS reagent solution
to the water sample.
Caution:
vial is filled up to the top!
6. Close the vial rightly with the cap and swirl gently several times to mix the
contents.
7. Place the vial in the sample chamber, making sure that the triangular
marks on the vial and instrument are aligned.
8. Pres [TEST] key.
The display shows:
9. Remove the vial from the sample chamber, empty the vial, rinse the vial
and cap several times and the fill the vial with exactly
10 ml Fluoride
standard
(concentration 1 mg/l F).
< Calibration >
191 Ca-hardness 2 T
Reset?
YES: 1
NO: 0
< Calibration >
1: M191 Ca-hardness 2
2: M191 reset 0 cali.
3: M170 Fluoride L
< Calibration >
170 Fluoride L
ZERO: deionised water
press ZERO
< Calibration >
170 Fluoride L
T1 accepted
T2 1 mg/l F
press TEST
< Calibration >
170 Fluoride L
T1: 0 mg/l F
press TEST
Содержание WSL25 Pro
Страница 1: ...AquaSafe WSL25 Pro Water Safety Laboratory Instruction Manual ...
Страница 7: ...Section 3 AquaSafe Micro biological Incubator Instruction Manual ...
Страница 25: ...Section 4 AquaSafe Membrane Filtration Instruction Manual ...
Страница 41: ...Section 5 HydroTest Photometer HT1000 ...
Страница 108: ...Section 6 HydroCheck HC1000 Multi Parameter Electrochemical Water Tester Instruction Manual ...
Страница 134: ...200 2000µS cm 700 1700µS cm 1413µS cm 2 20mS cm 7 17mS cm 12 88mS cm 20 200mS cm 70 170mS cm 111 8mS cm ...
Страница 141: ......
Страница 147: ...Zero oxygen solution is contaminated Replace the calibration solution ...
Страница 152: ...NOTES ...