Blank analysis:
•
Ensure that the 10ml vial is clean.
•
Fill the vial with 10ml of the water sample.
•
Fit the cap and tighten.
•
Place the vial in the sample chamber, making sure that the arrow marks on the instrument
and vial are aligned.
•
Press the zero key
•
Wait until the following is displayed
Sample preparation:
•
Add
one Nitrite LR tablet
straight from the foil to the water sample.
•
Crush the tablet using a clean stirring rod, until no large pieces are visible.
•
Close the vial tightly with the cap and swirl several times until the tablet is dissolved.
Analysis :
•
Place the vial in the sample chamber, making sure that the marks on the instrument and vial
are aligned.
•
Press the Test key
•
Wait for a reaction period of
10 minutes
•
After the reaction period, the measurement starts and the result is displayed in mg/l Nitrite.
LOD/Tolerance
Zero accepted
Prepare Test
Press TEST